Preparation
O
flowmeter
2
Prepare the flowmeter in accordance with its
specific Instructions for Use.
1
3
2
4
To mount the flowmeter on the compact rail:
1 Press the slider all the way down; hook the
flowmeter to the compact rail and release the
slider.
2 Screw on the O
connecting hose.
2
3 Insert the connector of the connecting hose into
the O
delivery socket and press it all the way in.
2
Nebuliser
Prepare the nebuliser in conformity with its
specific Instructions for Use.
4 Screw the nebuliser securely to the flowmeter.
5 Fit the spiral hose to the nebuliser port.
O
monitor
2
Prepare the O
monitor in conformity with its
2
specific Instructions for Use.
34
5
Fasten O
monitor, complete with its holder, to
2
the compact rail.
Place the O
sensor on the bed and feed the
2
cable through the hole on the front.
Plug the connector of the sensor into the
measuring unit.
O
supply via injector
2
When using the injector, the higher side panels
(230 mm) and the bed canopy must be used
(Page 30 and following pages).
6
6 Press locking lever on mounting flange and
push injector into flange.
Allow locking lever on injector to engage.
The injector is now locked.
Connect up hose.
Always monitor the O
using O
supply.
2
Fitting the hose bracket
Insert the hose bracket into the hole in the left side
of the head panel and tighten with knurled screw.
Instructions for use Babytherm 8004/8010
concentration when
2