Setting the Reference in Smart Remote Mode

1
Turn on the tester and remote and let them warm
up for 5 minutes.
2
On the main tester: set the following on Loss/Length
tab in Setup:
Set REMOTE END SETUP to Smart Remote.
Set THIS UNIT to Main.
3
For an OptiFiber smart remote: set the following on
Loss/Length tab in Setup:
Set REMOTE END SETUP to Smart Remote.
Set THIS UNIT to Remote.
4
Clean the tester's and source's OUTPUT connectors
and the reference test cord connectors.
5
On the main tester: press
Loss/Length Reference.
6
Make the connections shown on the SET REFERENCE
screen and in Figure 6-6; then press
6-18
F
; then select Set
t
.
7
Optional: From the TEST SETUP screen, you can enter
reference test cord lengths to meet TSB-140
reporting requirements:
L
Use
to highlight a reference test cord number,
t
press
; then enter a length. Press
are done.
C
8
Press
OK to leave the TEST SETUP screen.
W
Caution
Do not disconnect the reference test cords from
the tester's or remote's outputs after setting
the reference. If you do, you must set the
reference again to ensure valid measurements.
See "About Setting the Reference" on page 6-13 for
more information about referencing.