Fluke OF-500-10 Technical Reference Handbook page 186

Of-500 optifiber series
Hide thumbs Also See for OF-500-10:
Table of Contents

Advertisement

OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
Loss is negative.
The connections to the tester were disturbed after referencing.
The fiber ends were dirty during referencing.
There was a kink in a reference reference test cord during referencing.
The connectors were not properly aligned during referencing.
Some other problem caused a bad reference value.
If loss is negative, set the reference again and retest the cabling.
A known length of cable measures too long or too short.
The wrong fiber type is selected on the Cable tab in Setup.
The index of refraction needs adjustment. Change n on the Cable tab in Setup.
6-42
Table 6-10. Diagnosing Loss/Length Test Failures (cont.)