Operation

Use a machine which has been cleaned and
disinfected and is ready for operation.
Care, page 17
Preparation, page 20
Checking readiness for operation, page 28
In order to detect any malfunctions quickly, the following
parameters should be monitored during ventilation:
Check airway pressure, e.g. the PEAK and PEEP
parameters, on the pressure gauge or on the display
Check minute volume (set by means of "»Freq.« and
»V
«") on the display.
T

Using controlled ventilation IPPV

For ventilation frequencies of 5 to 40 1/min.
During IPPV, the minimum ventilation frequency is limited
to 5 1/min by Oxylog 2000.
1 Switch for ventilation modes set to IPPV.
To speed up settings, set the scale range to the same
colour on the »Freq.« and »V
the ventilation parameters are set as appropriate to the
patient group concerned, namely infants / children /
adults.
2 Set »Freq.« and »V
« knobs.
T
Body weight
Freq.
(kg)
1/min
Green range
30 to 40
for infants
(7.5 to 20)
Blue range
20 to 30
for children
(20 to 40)
Brown range
5 to 20
for adults
(over 40)
« knobs. This ensures that
T
V
T
litres
0.1 to 0.3
0.3 to 0.8
0.8 to 1.5
Operation
Using controlled ventilation IPPV
2
2
1
7