Dräger Polytron 5000 Instructions for Use WARNING Strictly follow the Instructions for Use. The user must fully understand and strictly observe the instructions. Use the product only for the purposes specified in the Intended use section of this document. Arrow.com.
Page 2
Arrow.com. Arrow.com. Downloaded from Downloaded from...
The instruments or components may not be modified in any manner. The use of faulty or incomplete parts is forbidden. The appropriate regulations must be observed at all times when carrying out repairs on these instruments or components. Dräger Polytron 5000 Arrow.com. Arrow.com. Arrow.com. Arrow.com.
Description Description Intended use The Dräger Polytron 5000 is a transmitter for the detection of Product overview carbon monoxide, hydrogen sulphide, oxygen, hydrogen, ammonia, chlorine, sulfur dioxide and nitrogen dioxide in The instrument is powered by 16 to 30 VDC. Gas ambient air.
Polytron 5000 rests exclusively with the end user; if in doubt 3 Direct current about placement, consult with Dräger application engineers. supply Polytron 5000 is a two-wire instrument powered by 16 to 30 V 4 Measured gas DC. Wiring terminals are located on the bottom of the 5 Membrane enclosure;...
Operation Installing the sensor Connect the wires as shown below. The Polytron 5000 sensor connector is keyed and can only be Secondary circuit intended to be supplied from an installed one way. Use only DrägerSensors in this instrument. isolating source.
‘sns’, ‘loc’, ‘1’, the instrument will switch to the measurement magnetic wand when the desired menu item is displayed. The mode, accepting the new sensor type. current value or status will flash to indicate a change to data Dräger Polytron 5000 Arrow.com. Arrow.com. Arrow.com.
This menu item is not active for the Polytron 5000. 3.6.3 Password The use of a password is optional with the Polytron 5000. A 3.6.10 A1 lat password consists of a 3-digit number from 000 to 999; a value This menu item is not active for the Polytron 5000.
Allow the gas to flow for at least three minutes before proceeding. 4. Scroll through software menu to Zero Adj and tap OK. The current zero value will be displayed. 5. Wait for the zero to stabilize. Dräger Polytron 5000 Arrow.com. Arrow.com. Arrow.com. Arrow.com.
Resolution: if FSD > 50 the resolution is 1 if FSD < 50 the resolution is 0.1 Voltage of power supply 1) Only to be used with Polytron 5000 General Purpose part number Operating voltage 16 to 30 VDC 4544165.
6809705 DrägerSensor EO 6810740 1) Only to be used with Polytron 5000 General Purpose part number 4544165. 2) Notice: Gases which react with moisture will not repeatedly pass through the wire mesh, if the transmitter is installed in a non- climate controlled environment.
Page 14
Order list Dräger Polytron 5000 Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from...
Page 15
Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Arrow.com. Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from Downloaded from...