Error Messages - Elcometer 224 Operating Instructions Manual

Model t surface profile gauge
Hide thumbs Also See for 224:
Table of Contents

Advertisement

Icon
- - -

Error Messages

Under certain conditions, the gauge will display error messages. The
cause of the error should be corrected before proceeding.
Message
#1 - Probe
#3 - Probe
No Probe
Probe Unsupported
Probe Changed
Unstable Reading External electromagnetic interference or gauge is faulty.
Value Too Large
Calibration Error
Bluetooth PC
Reply
Error - EEPROM
Write
Gauge Memory Full Batch memory full. Delete one or more batches
55
R
Description
Battery Life Indicator
Replace Batteries
User Selectable Feature: Switched Off
User Selectable Feature: Switched On
Beep Volume Setting
Measurement over range
Cause / Action
Probe to gauge communication failure. Integral Gauges:
Contact Elcometer. Separate Gauges: Remove probe and
refit. If error persists, contact Elcometer.
Internal Error - contact Elcometer.
No probe connected. Connect probe.
Wrong probe connected. Connect correct probe.
Probe has been changed during a calibration or whilst batching.
Isolate from electromagnetic interference. If error persists,
contact Elcometer.
Calculated value is too large to be displayed. Contact Elcometer.
Incorrect calibration calculation. Re-calibrate. If error persists,
contact Elcometer.
Bluetooth link working at limit of its range. PC is busy running
other applications. Interference. PC application doesn't send
replies. Move gauge closer to PC. Shutdown other
applications. Remove source of interference. If error persists,
contact Elcometer.
Gauge EEPROM communication error. If problem persists,
contact Elcometer.
Model
www.elcometer.com
BT
BT
BT
BT
BT
BT