Click ‘Update Gauge’ to install the latest firmware. The icon displayed depends on the measurement mode selected. If ‘Immediate’ mode is selected, no icon is displayed. See Section 4.4 ‘Selecting the Measurement Mode’ on page en-4 for further information. en-3 www.elcometer.com...
Page 5
4.4 SELECTING THE MEASUREMENT MODE The Elcometer 224 has a choice of three measurement modes; ‘Immediate’, ‘Custom’ and ‘Standards’ - see the table below for an explanation of the different modes available. The mode selected depends on which International Standard the user is working in accordance with.
Page 6
The value of the last reading taken is displayed on screen in grey, below the reading circles (b). When the last of the predefined number of readings has been taken, the mean/average, highest or lowest measurement (as selected by the user) is displayed (c). en-5 www.elcometer.com...
Menu/Setup/Screen Settings/Screen Timeout. Press any key or tap the gauge to awaken it. The gauge will switch off automatically after 5 minutes of inactivity. - - - indicates reading outside range of probe. www.elcometer.com en-6...
Calibration foils can be used to check the accuracy of the gauge across its measurement range. The Elcometer 224 Model T has a choice of calibration methods as detailed below. The Elcometer 224 Model B uses the Zero calibration method.
Page 9
When prompted, either place the probe on to the glass zero tile or on the centre of the foil resting on the glass tile, making sure the pin goes through the hole in the centre of the foil. 90° www.elcometer.com en-8...
To display the Run Chart of the last 20 readings: Press the Batch softkey Select “New Batch” or “Open Existing Batch” Press the Display softkey and select “Readings & Run Chart”. Red points signify a reading outside the batch’s limits (if set). en-9 www.elcometer.com...
Patents Select Statistics 51.5 Legal Notices µm Number of Readings ¡ Regulatory Escape Delete Mean ¡ Lowest Reading ¡ Highest Reading ¡ Calibration Standard Deviation ¡ Coefficient of Variation ¡ Pressing the Cal Softkey starts the calibration process www.elcometer.com en-10...
® USB. For more information on ElcoMaster visit www.elcometer.com. All Elcometer 224 gauge firmware can be upgraded to the latest versions, as they become available. Elcometer 224 B & T models ® can be upgraded by the User via ElcoMaster .
All Elcometer 224 Surface Profile Probes have an accuracy of ±5% or ±5μm (±0.2mil) and are supplied with a glass zero plate, calibration test foils; nominal values 125μm (5.0mils) & 508μm (20mils) and an Elcometer test certificate.
Bluetooth SIG Inc and licensed to Elcometer Limited. Elcometer 224 Model T: Made for iPhone 6 Plus, iPhone 6, iPhone 5s, iPhone 5c, iPhone 5, iPhone 4s, iPhone 4, iPad Air 2, iPad mini 3, iPad Air, iPad mini 2, iPad (3rd and 4th generation), iPad mini, iPad 2, and iPod touch (4th and 5th generation).
Page 16
TMA-0523 Issue 08 - Text with cover 23302...