13 ERROR MESSAGES

Under certain conditions the Covermeter will display error messages (Figure 21). These messages
are normally cleared by pressing any one of the softkeys. The cause of the error will be indicated
by the message and should be corrected before proceeding.
Error message
#1 - PROBE
#2 - PROBE
#3 - PROBE
#4 - PROBE
#5 - PROBE
PROBE
UNSUPPORTED
PROBE CHANGED
VALUE TOO
LARGE
CLOCK
LANGUAGE
MEMORY
* Contact Elcometer or your Elcometer Supplier to arrange return.
Figure 21. Example error message - search head not connected
Table 1: Error messages
Causes
Search head-to-gauge
communication failure.
Corrupt data output from
probe.
Internal error.
Search head power supply
fault.
Internal error.
Old Covermeter software does
not support new search head.
Probe has been changed and
is not compatible with current
batch.
Numerical error.
Internal error.
Software error.
Action to take
Remove search head and refit.
If error persists, return to Elcometer*.
Try new search head.
If error persists, return to Elcometer*.
Return to Elcometer*.
Remove probe. If error persists, the
Covermeter is faulty. If error clears, the
probe is faulty.
Return faulty part to Elcometer*.
Remove search head and refit.
If error persists, return to Elcometer*.
Return to Elcometer* for software
upgrade.
Change to probe used when batch was
created.
Switch Covermeter off then on again.
If error persists contact Elcometer.
Return to Elcometer*.
Return to Elcometer*.
R