Supply Voltage - HBM T10F Mounting Instructions

Torque flanges
Hide thumbs Also See for T10F:
Table of Contents

Advertisement

Electrical connection
Binder 723
6
1
5
7
2
3
4
Top view
1)
Distribution system for electrical energy with greater physical expansion (over several test
benches, for example) that may possibly also supply consumers with high nominal (rated)
currents.
46
Assignment connector 3
Voltage supply and voltage output signal.
Conn.
Assignment
Binder
Pin
1
Torque measurement signal (voltage output; 0 V
2
Supply voltage 0 V;
3
Supply voltage 18 V ... 30 V DC
Torque measurement signal (voltage output;
4
"10 V)
5
No function
6
Calibration signal trigger 5 V - 30 V
7
Calibration signal 0 V;
Shielding connected to housing ground
7.5

Supply voltage

The transducer must be operated with a separated
extra‐low voltage (18...30 V DC supply voltage), which
usually supplies one or more consumers within a test
bench.
Should the equipment be operated on a DC voltage
1)
network
, additional precautions must be taken to
discharge excess voltages.
The notes in this chapter relate to the standalone
operation of the T10F without HBM system solutions.
A0608-14.0 HBM: public
)
T10F