Nominal torque can be exceeded statically up to the limit
torque. If the nominal torque is exceeded, additional
irregular loading is not permissible. This includes
longitudinal forces, lateral forces and bending moments.
Limit values can be found in chapter 14 "Specifications",
on page 70.

10.1 Measuring dynamic torque

The torque flanges can be used to measure static and
dynamic torques. The following rule applies to the
measurement of dynamic torque:
S The T10F calibration made for static measurements is
also valid for dynamic torque measurements.
S The natural frequency f
measuring system depends on the moments of inertia
J
and J
1
T10F torsional stiffness.
Use the equation below to approximately determine the
natural frequency f
arrangement:
· 1
) 1
c
T
J
J
1
2
S The oscillation width (peak‐to‐peak) can be max. 160
% (for nominal (rated) torques 50 N@m=320 %, 10
kN@m=120%) of the nominal (rated) torque
designated for the T10F, even under alternating load.
The vibration bandwidth must fall within the loading
A0608-14.0 HBM: public
0
of the connected rotating masses and the
2
of the mechanical measuring
0
f
= natural frequency in Hz
0
J
J
= mass moment of inertia in kg⋅m
1,
2
c
= torsional stiffness in N⋅m/rad
T
Loading capacity
for the mechanical
2
59