Section 10

10.1 Weekly Maintenance

Oxidation Phase:
Check the oxygen cylinder/oxygen concentrator.
Check the oxygen pressure in the cylinder or the oxygen supply from the oxygen concentrator.
Check the oxygen pressure on the Oxygen Regulator inside the BioTector.
Check the Acid container and Acid Pump.
Check the Base container and Base Pump.
Check the HCl Water solution container and HCl Water Pump (if installed).
Check and confirm that the sample line is free and the Sample Pump is running correctly.
Check and confirm that the Circulation Pump is operating.
Check and confirm that all valves are operating.
Check the filter mats in the Fan and Vent housings.
Check the Bypass, Sample Out and Exhaust ports.
Liquid Phase:
Check the Oxidized Sample Catch-Pot/Cleaning Vessel.
Check Nitrogen Pump (N Pump).
Check the TN Cleaning container.
Check the DI Water container.
Check and confirm that all valves are operating.
Check the Zero Water (DI Water) used for the Zero Calibration.
Check the TN Drain and TN Bypass ports.