Ozone Test

WARNING
General overview of the operation of the ozone test:
Phase 1:
Install the tester according to information sheet T006, and start the test from the menu.
Phase 2:
The BioTector carries out a pressure test, to ensure that the system is leak tight.
Phase 3:
The ozone generator is switched on, and when the o-ring in the tester breaks, press the stop
test menu item.
Phase 4:
There is a purge period that purges any traces of ozone from the ozone tester, and the result
of the test is shown on the screen.
Phase 5:
The purge of the tester is complete, and the result remains on the screen.
Ozone Test, Phase 1:
O Z O N E
T E S T
1 < * S T A R T
T E S T
2
S T O P
T E S T
This menu enables the user to test the concentration of ozone generated by the BioTector.
1. Start Test. This starts the ozone test.
2. Stop Test. This stops the ozone test. It should be activated when the o-ring in the tester breaks, or at any
time to stop the ozone test.
Ozone Test, Phase 2:
P R E S S U R E
T E S T
T I M E
M F C
S E T P O I N T
M F C
F L O W
S T A T U S
P R E S S
E S C
This menu enables the user to monitor the progress of the ozone test. To abort the ozone test, press the
ESCAPE key on the keyboard.
The ozone test uses the procedure described in information sheet "T006,
Procedure to check the ozone level". The user must read and understand the
processes described in this sheet, and have all the correct parts listed before
carrying out the test.
3 5 s
4 0 . 0 l / h
2 2 . 0 l / H
T E S T I N G
T O
A B O R T
0 9 : 1 7 : 2 8
1 2 - 0 9 - 0 2
0 9 : 1 7 : 2 8
1 2 - 0 9 - 0 2
T H E
T E S T