Table 5 Various BioTector Dimensions
BioTector TOC TN Analyzer
BioTector TOC TN Analyzer
with integral oxygen
concentrator
BioTector TOC TN Analyzer
with extended lower
enclosure
BioTector TOC TN Analyzer
with integral oxygen
concentrator and extended
lower enclosure
Figure 8 below illustrates the BioTector door clearance dimensions.
Figure 8 BioTector Door Clearance Dimensions
When BioTector is being mounted on a wall or a stand, the support has to be strong enough to carry
typically four times of the weight of BioTector (~400 kg).
The BioTector should be lifted applying a safe method in accordance with local regulations.
The minimum size of the bolts used to hold the BioTector in place should be M8.
Dimension A
Dimension B
500mm
750mm
750mm
750mm
500mm
1000mm
750mm
1000mm
Dimension C
Dimension D
1025mm
1250mm
1150mm
1500mm
1275mm
1500mm
1400mm
1750mm