Hach BioTector B7000 User Manual page 140

Online toc tn analyzer
Table of Contents

Advertisement

FAULT
18_LIQUID LEAK DET
19_DCP LIQ. LEAK DET
20_NO REAGENTS
(can also be programmed as
a warning or notification
condition)
CONDITION
BioTector's cabinet liquid leak detector
is activated.
BioTector's DCP module liquid leak
detector is activated.
BioTector has calculated that the
reagent containers should be empty.
CAUSE/REMEDY
Check for a liquid leak in the
BioTector.
Check BioTector liquid leak
detector.
Check for a liquid leak in the
BioTector DCP module.
Check DCP module liquid leak
detector.
Change reagents, and reset
reagent monitor in Install New
Reagents menu.
Confirm reagent volume/container
size input values are correct in
Reagents Monitor menu.
Page 140

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents