Hach BioTector B7000 User Manual page 21

Online toc tn analyzer
Table of Contents

Advertisement

Reagent Status
B I O T E C T O R
A C I D
M O N I T O R
B A S E
M O N I T O R
H C l
W A T E R
T N
C L E A N I N G
D I
W A T E R
N O T E :
If the Reagent Status screen has been activated, the estimated number of days left for each reagent type is
shown on the display.
If the reagents run low, a LOW REAGENTS fault is activated. This fault has to be cleared by resetting the
reagent level in the Install New Reagents menu.
Note that the LOW REAGENTS fault can be set as a warning (where the common fault relay will activate) or a
notification, in which case a special programmable relay is required to signal the LOW REAGENTS condition.
Select Level
S E L E C T
L E V E L
1 <
O P E R A T I O N
2
C A L I B R A T I O N
3
M A I N T E N A N C E
This screen allows the user to access the operation, calibration and maintenance menus.
1. Operation. This menu gives access to the basic operation of the BioTector and allows access to the
archives. The level can be password protected using the Password menu.
2. Calibration. This menu allows the user to run zero and span calibration cycles. The level can be
password protected using the Password menu.
3. Maintenance. This menu allows the user to test the individual components of the BioTector for
diagnostics, to download data, to program the software functions and to program the system specific
settings in the BioTector. The sub menus in this level can be password protected using the Password
menu.
R U N N I N G
M O N I T O R
M O N I T O R
M O N I T O R
D A Y S
L E F T
I S
B A S E D
S Y S T E M
0 9 : 1 7 : 2 8
1 2 - 0 9 - 0 2
2 5 . 0 l
~
3 5
2 5 . 0 l
~
3 5
1 0 . 0 l
~
3 8
2 5 . 0 l
~
1 5 0 D A Y S
2 5 . 0 l
~
6 0
A N
E S T I M A T E
C U R R E N T
U S E
0 9 : 1 7 : 2 8
1 2 - 0 9 - 0 2
D A Y S
D A Y S
D A Y S
D A Y S
Page 21