OXIDATION PROGRAM 2 TC
BASE IN 1
SAMPLE IN
BASE IN 2
TOC ACID
OXIDATION PROGRAM 2 VOC
OXIDATION PROGRAM 2 TC - TIC
TC SAMPLE IN
TIC SAMPLE IN
TIC ACID
TOC ACID
OXIDATION PROGRAM 3
OXIDATION PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3
TN
RANGE CHANGE 3-2
TN
RANGE CHANGE 3-1
TN
12p
BASE IN 1 specifies the quantity of base reagent (in pulses) to
be injected into the reactor for the Pre Oxidation phase in TC
mode at range 2.
5p , M-V
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
3p
BASE IN 2 identify the quantity of base reagent (in pulses) to be
injected into the reactor for the VOC Oxidation phase in TC
mode at range 2.
15p
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase in TC mode at
range 2.
All system parameters for OXIDATION PROGRAM 2 VOC
analysis mode are covered in Oxidation Program 2 TC and
Oxidation Program 2 TIC + TOC above.
5p , M-V
TC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TC reactor at range 2 in TC - TIC systems.
5p , P-V
TIC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TIC reactor at range 2 in TC - TIC systems.
TIC ACID determines the amount of acid reagent (in pulses) to
3p
be injected into the TIC reactor at range 2 in TC - TIC systems.
12p
TOC ACID states the amount of acid reagent (in pulses) to be
injected into the TC reactor at range 2 in TC - TIC systems.
RANGE CHANGE 2-3 defines the concentration level for the
150.0 mgC/l
75.0mgN/l
BioTector to change its range automatically from range 2 to
range 3.
The RANGE CHANGE 2-3 is typically the top concentration
point of TOC/TN RANGE 2 in System Range Data menu by
default.
RANGE CHANGE 3-2 specifies the concentration level for the
120.0mgC/l
60.0mgN/l
BioTector to change its range automatically from range 3 to
range 2.
The RANGE CHANGE 3-2 is typically 20% less of the top
concentration point of TOC/TN RANGE 2 in System Range Data
menu by default.
40.0mgC/l
RANGE CHANGE 3-1 defines the concentration level for the
20.0mgN/l
BioTector to change its range automatically from range 3 to
range 1.
The RANGE CHANGE 3-1 is typically 20% less of the top
concentration point of TOC/TN RANGE 1 in System Range Data
menu by default.
Page 119