OXIDATION PROGRAM 2 TC
BASE IN 1
SAMPLE IN
BASE IN 2
TOC ACID
OXIDATION PROGRAM 2 VOC
OXIDATION PROGRAM 2 TC - TIC
TC SAMPLE IN
TIC SAMPLE IN
TIC ACID
TOC ACID
OXIDATION PROGRAM 3
OXIDATION PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3
TN
RANGE CHANGE 3-2
TN
RANGE CHANGE 3-1
TN
12p
 BASE IN 1 specifies the quantity of base reagent (in pulses) to
be injected into the reactor for the Pre Oxidation phase in TC
mode at range 2.
5p , M-V
 SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
3p
 BASE IN 2 identify the quantity of base reagent (in pulses) to be
injected into the reactor for the VOC Oxidation phase in TC
mode at range 2.
15p
 TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase in TC mode at
range 2.
 All system parameters for OXIDATION PROGRAM 2 VOC
analysis mode are covered in Oxidation Program 2 TC and
Oxidation Program 2 TIC + TOC above.
5p , M-V
 TC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TC reactor at range 2 in TC - TIC systems.
5p , P-V
 TIC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TIC reactor at range 2 in TC - TIC systems.
 TIC ACID determines the amount of acid reagent (in pulses) to
3p
be injected into the TIC reactor at range 2 in TC - TIC systems.
12p
 TOC ACID states the amount of acid reagent (in pulses) to be
injected into the TC reactor at range 2 in TC - TIC systems.
 RANGE CHANGE 2-3 defines the concentration level for the
150.0 mgC/l
75.0mgN/l
BioTector to change its range automatically from range 2 to
range 3.
 The RANGE CHANGE 2-3 is typically the top concentration
point of TOC/TN RANGE 2 in System Range Data menu by
default.
 RANGE CHANGE 3-2 specifies the concentration level for the
120.0mgC/l
60.0mgN/l
BioTector to change its range automatically from range 3 to
range 2.
 The RANGE CHANGE 3-2 is typically 20% less of the top
concentration point of TOC/TN RANGE 2 in System Range Data
menu by default.
40.0mgC/l
 RANGE CHANGE 3-1 defines the concentration level for the
20.0mgN/l
BioTector to change its range automatically from range 3 to
range 1.
 The RANGE CHANGE 3-1 is typically 20% less of the top
concentration point of TOC/TN RANGE 1 in System Range Data
menu by default.
Page 119