TIC CHECK
TOC CHECK

Result Integration

RESULT INTEGRATION
TOC INTEGRATION
TOC INT. LIMITS
TN INTEGRATION
TN INT. LIMITS
25ppmCO2
 TIC CHECK (25ppm CO
 TOC CHECK (100ppm CO
100ppmCO2
TOC+BASE OXID
 RESULT INTEGRATION determines the reaction point where
3
 TOC INTEGRATION (3 by default) defines the number of
6% , 5.0
 TOC INTEGRATION LIMITS controls the averaging function
 TN INTEGRATION (3 by default) defines the number of reaction
3
10% , 5.0
 TN INTEGRATION LIMITS controls the averaging function
by default) represents the CO
2
point in the TIC phase. If the CO
check point at the end of the TIC phase, then the system
automatically extends the TIC SPARGE TIME by 1 second and
checks the CO
level again. If the TIC level does not drop below
2
the check point at the end of the maximum 300 seconds, a
"50_TIC OVERFLOW" warning is generated.
check point in the TOC phase. If the CO
programmed check point at the end of the TOC Oxidation
section of the TOC phase, then the system automatically
extends the TOC SPARGE TIME and TOC OXIDATION time by
1 second and checks the CO
not drop below the check point at the end of the maximum 300
seconds, a "51_TOC OVERFLOW" warning is generated.
TOC result integration is started. With the default setting of
"TOC+BASE OXID", the result integration starts with the Base
Oxidation phase and completes at the end of the TOC phase. If
the setting is "TOC" only, the result integration is carried out only
during the TOC phase.
reaction results to be averaged to get the actual TOC result.
defined in TOC INTEGRATION above. The first parameter "6%"
defines the % band variation and the second parameter "5.00"
defines the absolute variation in mgC/l. If the reaction result is
outside the specified bands (TOC INTEGRATION LIMITS), then
the reaction result averaging is omitted. In other words, the
result which is outside the INTEGRATION LIMITS is displayed at
the end of the reaction without averaging. If the result is within
the TOC INTEGRATION LIMITS, the averaging function is
continued and the programmed number of reaction results
(defined in TOC INTEGRATION above) is averaged.
results to be averaged to get the actual TN results.
defined in TN INTEGRATION above. The first parameter "10%"
defines the % band variation and the second parameter "5.00"
defines the absolute variation (in mgN/l). If the reaction result is
outside the specified bands (TN INTEGRATION LIMITS), then
the reaction result averaging is omitted. In other words, the
result which is outside the INTEGRATION LIMITS is displayed at
the end of the reaction without averaging. If the result is within
the TN INTEGRATION LIMITS, the averaging function is
continued and the programmed number of reaction results
(defined in TN INTEGRATION above) is averaged.
level is above the programmed
2
by default) represents the CO
2
level is above the
2
level again. If the TOC level does
2
check
2
2