Minimum Software Requirements; Reportable Range Limitations - Abbott i-STAT 1 System Manual

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

Note: When running controls, if the handheld is customized for User Defined Reportable
Ranges and Auto Pass/Fail detection, the handheld will display the control result based upon
the customized Reportable Range setting, but will base the Auto Pass/Fail determination on the
measured value, irrespective of the Reportable Range settings.
3. Narrowing the reportable range of certain tests may affect the presentation of other dependent test
results. In the table below, if any of the results in the first column are outside the reportable range,
the dependent test results listed in the second column will be suppressed (displayed as <>).
Test (Outside the Reportable Range)
Na
Hct
PCO
pH
HCO
Cl
K
TCO
PO
PT/INR

MINIMUM SOFTWARE REQUIREMENTS

The minimum handheld and data management requirements for use of the Custom Reportable Range
feature are summarized as follows:
System Component
i-STAT 1 handheld
Central Data Station Version 5.x
i-STAT/DE
2

Reportable Range Limitations

2
3
2
2
Art: 730009-00B
Dependent Suppressed Tests (displayed as <>)
K, Cl, BUN, Anion Gap, Hgb, Hct
Cl, BUN, Anion Gap, Hgb
TCO
, Anion Gap, Base Excess, HCO
2
TCO
, Anion Gap, Base Excess, HCO
2
TCO
, Anion Gap, Base Excess, sO
2
Anion Gap
Anion Gap
Anion Gap
sO
PT WBT
Minimum Software Requirements
≥ JAMS132
≥ Version 5.26a
≥ Version 2.3
, sO
3
2
, sO
3
2
2
2
Rev. Date: 09-Jan-12