Printing Many Results; What Is Printed - Abbott i-STAT 1 System Manual

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

Printing Many
Select 2 – Data Review from the Administration Menu on the analyzer, then
Results
select 7 – List. Use the arrow keys to page up and down through the pages of
stored results. Press the numbered key for each test record to be printed. To
deselect a record, press that numbered key again. When all test records have
been selected, align the printer and analyzer according to the directions under
Printing Directly from an Analyzer or place the printer in a Downloader or
Downloader/Recharger according to the directions above, and press the Print
key.

What is Printed

Name of Test
Sample ID
Results
At Patient Temperature
Sample Type
Free Fields
Time and Date
Operator ID
Lot Number
Serial
Version
CLEW
Caution
7 - 4
Use only a rechargeable battery pack purchased from Abbott Point of
Care. Rechargeable battery packs not recommended by or purchased
from Abbott Point of Care may be susceptible to overheating and could
lead to a potential fire or burn hazard.
Use power supply provided with printer.
Do not operate the printer without paper.
Do not allow the power supply to become a trip hazard.
Do not disturb the analyzer or printer until printing is complete since
this will interrupt the printout. If printing is interrupted, realign the
printer and analyzer or replace the analyzer in the Downloader to resume
printing. Note: If significant time has elapsed, some results may be
missing from the printout. Reprint the results.
If printed results appear inconsistent with a patient's clinical assessment,
verify that the printed results match the data in the analyzer. If the
results do match, the patient sample should be retested using another
cartridge. If they do not match, reprint the results. If the reprint still
does not match the analyzer data, the printer requires service and the
printed results must not be used.
Art: 714369-00H
i-STAT cartridge type
Patient ID or type of quality test and lot
number of solution tested
Results are printed with units as well as
flags and comment codes if applicable.
If the patient's temperature was entered
on the Chart Page, a second set of results
is displayed for blood gases at the patient's
temperature.
Sample type selected from Chart Page
when sample is patient or proficiency test
Information entered into the free fields on
the Chart Page when sample is patient or
proficiency test
Time and Date when test was performed
Operator ID
Lot number of cartridge
Serial number of the analyzer
Analyzer application software
Standardization software
Rev. Date: 03-Aug-12