Verifying Software Update

1. 1 7
Run the Electronic Simulator in the Handheld.
When the simulator finishes, PASS should be displayed.
Congratulations. The process for updating
the first i-STAT 1 Handheld is complete.
Review the options below for additional instructions.
If there are no additional Handhelds to update, the process is complete:
Click the
Close all other open boxes
Confirm all messages
If there are additional Handhelds to update via JammLite:
Click Close
Click Update
Repeat Steps
If there are additional handhelds to update with the Handheld-to-Handheld process:
Go to
If your facility uses Central Data Station:
You must proceed to
Art: 725708-00E
Rev. Date 07-FEB-12
X
button in the upper right corner of the software screen
1. 1 4
1. 1 7
through
2.
on page 11
3.
i-STAT 1 JammLite Utility
If PASS is not displayed,
re-run the Electronic Simulator.
If the repeated Electronic
Simulator attempt fails, please
contact APOC Technical Support.
For additional information on
running the electronic simulator,
please see:
• Section 14 of the i-STAT 1
System Manual, or
• The Introduction and Start-up
section of the i-STAT System
Manual for Waived Tests
on page 16.