i-Stat analyzer
and Philips BaM
CleW Windows
i-Stat 1 Software
Window
Rev. Date: 07/18/06
A new CLEW is added to the window via the software update process twice a
year. If the update CLeWs feature is active, the Default customization profile
must be updated after each new CLEW is added. Click the new CLEW and click
OK. Note that there are separate CLEWs for the i-STAT analyzer and the Philips
Blood Analysis Module.
note: Before changing to a new CLEW ensure that all analyzers have been
updated to a compatible application software version.
A new JAMS application software is added to the window via the software
update process twice a year. Click the new JAMS and click OK. Selecting the
new JAMS aloows users the option of updating their i-STAT 1 analyzers remotely
or locally using the Customization Workspace. Please see the Technical Bulletin
"Updating Analyzer Software" for full details on performing this procedure.
Checking "Disable Software Updates" disables this feature.
Art: 714384-00D