Continuity Measurement - AEMC 6536 User Manual

Megohmmeter
Table of Contents

Advertisement

Measurement
Range
Resolution
Accuracy
Typical test voltage vs load curve
The voltage as a function of the measured resistance is illustrated below:
The range of operation per IEC 61557 is from 100kΩ to 2GΩ (see § 3.3.4).

3.2.3 Continuity Measurement

Specific reference condition: Inductance in series with the resistance = zero.
Measurement Range (without
compensation of the leads)
Resolution
Accuracy
Test Current
Open Voltage
*In the case of incorrect compensation of the leads, the instrument allows display
of negative values, down to -0.05Ω at 200A and -0.5Ω at 20mA.
Test Current
200mA range: 200mA (0mA + 20mA)
20mA range: 20mA ± 5mA
Measurement Range
Megohmmeter Model 6536
Shop for Data Logging products online at:
0.01 to
39.99µA
10nA
± (10% R + 3 ct)
0.00* to 10.00Ω
10mΩ
200mA
DataLoggerStore
www.
40.0 to
399.9µA
100nA
0.0* to 100.0Ω
± (2% + 2 ct)
≥ 6V
0 to 250mA
.ca
0.400 to
2.000mA
1µA
100mΩ
20mA
31