Continuity Measurement - AEMC 502 User Manual

Trms clamp-on meter
Hide thumbs Also See for 502:
Table of Contents

Advertisement

4.5
• Turn the rotary range switch to the
range.
• Insert red test lead to the red "+" input jack
and the black lead to the black "COM" input
jack.
• Bring the test probe tips into contact with the
sample under test.
• If the resistance is less than 40Ω, the beeper
emits a continuous sound.
TRMS Clamp-on Meter Model 502
ts Online at:

Continuity Measurement

WARNING: Immediately unclamp the
meter from the conductor under test if
overload "
WARNING: When testing continuity,
make sure that there is no power in
the tested sample or circuit (dead cir-
cuit). This may be checked by using
the voltage functions.
MODEL 502
®
TRMS CLAMP METER
I N S T R U M E N T S
600V CAT II
400A
300V CAT III
HOLD
H
GlobalTestSupply
www.
" is displayed.
I N S T R U M E N T S
600V CAT II
300V CAT III
HOLD
H
MODEL 502
®
TRMS CLAMP METER
400A
13
.com
sales@Glo

Advertisement

Table of Contents
loading

Table of Contents