To turn OFF the alarm buzzer while it is sounding, press the HOLD button. To
deactivate an active alarm function, press the
2.1.2.2 Setting an alarm threshold
While
the alarm threshold by pressing the button (except during insulation
measurements). For each testing mode, there are three pre-defined threshold
values:
■ Continuity: < 2Ω, <1Ω and <0.5Ω
■ Resistance: >50kΩ, >100kΩ and >200kΩ
■ Insulation: <10kΩ, <50kΩ and <100kΩ
In each measurement mode, the third threshold can be replaced by a user-
defined value. To do this:
1. Press the ► button while the threshold value is displayed.
2. The > symbol starts blinking; you can change it to < by pressing the
button. This symbol indicates the direction of the alarm threshold: < for a
low threshold and > for a high threshold.
3. To change the threshold setting, press the ► button to navigate to the
first digit, and then use the button to change its value.
4. Use the ► and buttons to select and change the other digits in the
threshold value, as well as the units of measurement.
5. When finished setting the threshold, press the ► button to validate the
setting.