Vega VEGAPULS 63 Operating Instructions Manual page 40

Radar sensor for continuous level measurement of liquids
Hide thumbs Also See for VEGAPULS 63:
Table of Contents

Advertisement

6 Set up the sensor with the display and adjustment module
40
– Entering the tube inside diameter takes the running time shift
into consideration
– Echo detection sensitivity reduced
Bypass:
Medium speed:
– Fast up to slow filling with short up to long bypass tube possible
– Often the level is hold via a control facility
Vessel:
– Lateral outlets and inlets
– Joins like flanges, weld joints
– Shifting of the running time in the tube
Process/measurement conditions:
– Condensation
– Buildup
– Separation of oil and water possible
– Overfilling into the antenna possible
Properties, sensor:
– Measurement speed optimized through little averaging
– Entering the tube inside diameter takes the running time shift
into consideration
– Echo detection sensitivity reduced
– False signal suppression recommended
Plastic tank:
Vessel:
– Instrument fix mounted or built in
– Measurement through the vessel top, if appropriate to the
application
– With empty vessel, the measurement can go through the bot-
tom
Process/measurement conditions:
– Condensation on the plastic ceiling
– In outdoor facilities, water and snow on vessel top possible
Properties, sensor:
– False signals outside the vessel are not taken into consideration
– False signal suppression recommended
Transportable plastic tank:
Vessel:
– Material and thickness different
– Measurement through the vessel top
Process/measurement conditions:
– Measured value jump with vessel change
Properties, sensor:
– Quick adaptation to changing reflection conditions due to ves-
sel change
– False signal suppression required
Open water (gauge measurement):
Rate of level change: slow level change
Process/measurement conditions:
– Large distance from sensor to water surface
VEGAPULS 63 • Modbus and Levelmaster protocol

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vegapuls 62

Table of Contents