5 Connecting to power supply and bus system
Electronics compartment
Terminal compartment
30
( )
Fig. 18: Electronics compartment - double chamber housing
1
Internal connection to the terminal compartment
2
For display and adjustment module or interface adapter
Information:
The connection of an external display and adjustment unit is not pos-
sible with this double chamber housing.
( )
4
Fig. 19: Terminal compartment
1
USB interface
2 Slide switch for integrated termination resistor (120 Ω)
3
Modbus signal
4
Voltage supply
Terminal
Function
1
Voltage supply
2
Voltage supply
3
Modbus signal D0
4
Modbus signal D1
5
Function ground when installing ac-
cording to CSA (Canadian Standards
Association)
4...20mA
+
1
2
5
6
(-)
1
1
USB
MODBUS
power supply
+
1
2
off on
3 4
(-)
VEGAPULS 63 • Modbus and Levelmaster protocol
2
7 8
1
2
5
3
Polarity
+
-
+
-