4 Mounting
26
Constructional requirements on the bypass pipe:
Material metal, smooth inner surface
In case of an extremely rough tube inner surface, use an inserted
tube (tube in tube) or a radar sensor with tube antenna
Flanges are welded to the tube according to the orientation of the
polarisation
Gap size with junctions ≤ 0.1 mm, for example, when using a ball
valve or intermediate flanges with single pipe sections
The antenna diameter of the sensor should correspond to the
inner diameter of the tube
Diameter should be constant over the complete length
VEGAPULS 63 • Modbus and Levelmaster protocol