Display/Alarms; Table 7: Display/Alarms Problems - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Troubleshooting

Display/Alarms

46
Table 7 on page 46 lists symptoms of problems relating to non-functioning
displays and audible tones or alarms, and recommended actions. If the action
requires replacement of a PCB or module, refer to Disassembly Guide on
page 51.

Table 7: Display/Alarms Problems

Symptom
Display values are
missing or erratic.
Display pixels do not
light.
Alarm sounds for no
apparent reason.
Alarm does not sound.
Recommended Action
If the sensor is connected, replace the pulse
oximetry cable.
If the condition persists, replace the sensor.
If the condition still persists, replace the User
Interface printed circuit board. See User Interface
PCB Removal/Replacement on page 64.
Check the connection between the User Interface
PCB and the Display PCB.
If the condition does not change, replace the
Display PCB. See Display PCB Removal/
Replacement on page 62.
If the condition still persists, replace the User
Interface PCB. See User Interface PCB Removal/
Replacement on page 64.
Moisture or spilled liquids can cause an alarm to
sound. Allow the monitor to dry thoroughly
before using.
If the condition persists, replace the User
Interface PCB. See User Interface PCB Removal/
Replacement on page 64.
Check alarm silence status.
Check speaker connection.
Replace the speaker as described in Alarm
Speaker Removal/Replacement on page 66.
If the condition persists, replace the User
Interface PCB. See User Interface PCB Removal/
Replacement on page 64.