Satseconds Alarm Management; Reads Through Motion; Oximax Technology; Figure 25: Oxyhemoglobin Dissociation Curve - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

SatSeconds
Alarm
Management
Reads Through
Motion
O
M
XI
AX
Technology
N-595

Figure 25: Oxyhemoglobin Dissociation Curve

The N-595 utilizes Nellcor SatSeconds alarm management technique.
SatSeconds is a function of the software within the N-595. With the
SatSeconds technique, upper and lower alarm limits are set in the same way as
traditional alarm management. The clinician also sets a SatSeconds limit that
allows monitoring of %SpO
of time before an audible alarm sounds. Refer to the N-595 Operator's manual
for managing SatSeconds.
The N-595 takes advantage of increased micro processing power with
advanced mathematical algorithms. Oxismart
processing allows the N-595 to read through challenging motion conditions to
deliver accurate saturation and pulse rate values. For a definition of motion, as
applicable to the N-595, contact Nellcor's Technical Services Department.
The N-595 pulse oximeter is designed to use Nellcor brand sensors containing
O
M
technology. These sensors can be identified by the deep blue color of
XI
AX
their plug. All O
M
-compatible sensors contain a memory chip carrying
XI
AX
information about the sensor which the oximeter needs for correct operation,
including the sensor's calibration data, model type, troubleshooting codes,
and error detection data. This unique oximetry architecture enables several
new features with the N-595.
below the selected lower alarm limit for a period
2
®
XL advanced signal
Technical Discussion