Trend Data Printout (Ascii Mode); Table 19: Operating Status Codes - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Data Port Interface Protocol
Trend Data
Printout (ASCII
Mode)
104

Table 19: Operating Status Codes

Code
AO
Alarm Off
AS
Alarm Silence
BU
Battery in Use
LB
Low Battery
LM
Loss of Pulse with Motion
LP
Loss of Pulse
MO
MOtion
PH
Pulse Rate Upper Limit Alarm
PL
Pulse Rate Lower Limit Alarm
PS
Pulse Search
SD
Sensor Disconnect
SH
Saturation Upper Limit Alarm
SL
Saturation Lower Limit Alarm
- - -
No Data Available
*
Alarm Parameter Being Violated
Note:
A Sensor Disconnect will also cause three dashes (- - -) to be
displayed in the patient data section of the printout.
The format of data displayed when a trend printout is requested is similar to
that of the real-time data. The only differences are that "TREND" is displayed
in the top row instead of the "CRC:XXXX" software verification number, and
there is no "Status" column (Figure 22 on page 105).
Readings are displayed in 4-second intervals. The values on each row are an
average for the 4-second period.
At the end of the printout, an "Output Complete" line indicates that the
transmission was successful. If the "Output Complete" line is not present, the
data should be considered invalid.
Meaning