Table 16: Recommended Separation Distances - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Specifications
84

Table 16: Recommended Separation Distances

Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the N-595 (IEC 60601-1-2)
Frequency of
150 kHz to
Transmitter
Equation
d = 1.2√P
Rated Maximum
Separation
Output Power of
Distance in
Transmitter in
Watts
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the separation
distance can be estimated using the equation in the corresponding column, where
P is the maximum output [power rating of the transmitter in watts (W)] according
to the transmitter manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
80 MHz to
80 MHz
800 MHz
d = 1.2√P
Separation
Distance in
Meters
Meters
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
800 MHz to
2.5 GHz
d = 2.3√P
Separation
Distance in
Meters
0.23
0.73
2.3
7.3
23

Hide quick links: