Battery
Replacement

Removal

N-595
1. Follow the procedure in paragraphs Prior to Disassembly on page 52, and
Monitor Disassembly on page 53.
2. Remove the two screws from the battery bracket and lift the battery out of
the bottom case as shown in Figure 10 on page 55.
3. Be sure to note the polarity of the leads. Use needle-nose pliers to
disconnect the leads from the battery.

Figure 10: Removing the Battery

4. The lead-acid battery is recyclable. Do not dispose of the battery by
placing it in the regular trash. Dispose of the battery in accordance with
local guidelines or return it to Nellcor's Technical Services for disposal.