Nellcor OXIMAX N-595 Service Manual page 49

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

N-595
2. Set desired parameters to the institutional values. Refer to the N-595
operator's manual for the procedures to set the values.
3. Simultaneously press the LIGHT softkey and the CONTRAST button
until the menu bar changes to the softkey headings shown below.
4. Press the PARAM softkey.
5. Press the SAVE softkey.
6. Press the YES softkey. The monitor will sound three beeps indicating that
defaults have been saved.
Power-On Settings and Service Functions
41

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents