Power-On Settings And Service Functions; Introduction; Power-On Settings; Factory Default Settings - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

P o w e r - O n S e t t i n g s a n d S e r v i c e
F u n c t i o n s

Introduction

Power-On
Settings
Factory Default
Settings
Neonate Default
Settings
N-595
This section discusses how to reconfigure power-on default values and access
the service functions.
The following paragraphs describe how to change power-on default settings.
By using softkeys as shown in Figure 1 on page 3, the user can change alarm
limits, the type of display, baud rate, time and date, and trends to view.
Some values cannot be saved as power-on default values. An SpO
alarm limit less than 80 will not be saved as a power-on default. Audible
Alarm Off will not be accepted as a power-on default. An attempt to save
either of these values as default will result in an invalid tone. These limits can
be adjusted lower for the current patient, but they will be lost when the
instrument is turned off.
A decimal point is added to the right of a display when the alarm limit for that
display has been changed to a value that is not a power-on default value. If the
new value is saved as a power-on default value, the decimal point will be
removed. By using the service functions, changes can be saved as power-on
default values.
Factory default settings are divided into two groups, adult and neonate.
Default settings may be changed to institutional default setting , refer to
Setting Institutional Defaults (Sample), page 40.

Table 2: Neonate Alarm Limit Factory Defaults

Monitoring Mode
Note: Bold entries are different than adult default settings.
%SpO
Lower Alarm Limit
2
%SpO
Upper Alarm Limit
2
Alarm Silence Duration
Setting
80%
95%
60 seconds
lower
2
29