9. Press the SELECT softkey three times. Verify that the monitor emits a
beep after each keystroke. The Pulse upper alarm limit should be "170"
and should be boxed.
10. Press and hold the ADJUST DOWN button.
11. Verify that the minimum displayed value is "41" for the BPM upper alarm
limit.
12. Press the EXIT softkey.
13. Press the LIMITS softkey.
14. Press the SELECT softkey four times. Verify that the pulse rate lower
alarm limit display indicates an alarm limit of "40" and is boxed.
15. Press and hold the ADJUST DOWN button. Verify that the boxed pulse
rate lower alarm limit display reduces to a minimum of "30."
16. Press and hold the ADJUST UP button and verify that the boxed pulse
rate lower alarm limit display cannot be adjusted above the pulse rate
upper alarm limit of "40."
17. Press the EXIT softkey.
Performance Verification
13