Display/Alarms
O
M
N-600x Service Manual
XI
AX
Table 9 lists symptoms of problems relating to non-functioning displays and
audible tones or alarms, and recommended actions. If the action requires
replacement of a PCB or module, refer to Disassembling the Monitor on page 85.
Table 9: Display/Alarms Problems
Symptom
Display values are
missing or erratic.
Display pixels do not
light.
Alarm sounds for no
apparent reason.
Alarm does not sound.
Primary speaker fails
and a slow, high-pitched
tone sounds.
Recommended Action
O
M
•
If the
sensor is connected, replace the
XI
AX
pulse oximetry cable.
•
If the condition continues, replace the
sensor.
•
If the condition persists, replace the Main Board
printed circuit board. See Main Board PCB
Replacement on page 97.
•
Check the connection between the Main Board
PCB and the Display PCB.
•
If the condition continues, replace the Display
PCB. See Display PCB Replacement on page 95.
•
If the condition persists, replace the Main Board
PCB. See Main Board PCB Replacement on
page 97.
•
Moisture or spilled liquids can cause an alarm to
sound. Allow the monitor to dry thoroughly before
using.
•
If the condition persists, replace the Main Board
PCB. See Main Board PCB Replacement on
page 97.
•
Check the alarm silence status.
•
Check the speaker connection.
•
Replace the speaker as described in Alarm
Speaker Replacement on page 99.
•
If the condition persists, replace the Main Board
PCB. See Main Board PCB Replacement on
page 97.
•
Press the HELP/CONTRAST button and follow
the on-screen messages.
Troubleshooting
O
M
XI
AX
67