Nurse Call Setup

N-595
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the LANG softkey.
5. Use the ADJUST UP and ADJUST DOWN buttons to select the desired
language.
6. Press the EXIT softkey to save the language setting.
The voltage polarity for the Nurse Call, available at pins 11 and 5, can be
selected through the softkeys. NORM + sets the voltage to +5 VDC to +12
VDC and NORM - sets the voltage to -5 VDC to -12 VDC when there is no
audible alarm. When an audible alarm occurs, these voltages switch polarity.
This signal is available only if the instrument is operating on AC power. For
more information, see Nurse call page 105.
1. Turn on the N-595 by pressing the POWER ON/OFF button.
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
Data Port Interface Protocol
93