Specifications
82
Table 14: Electromagnetic Immunity (IEC 60601-1-2)
The N-595 is suitable for use in the specified electromagnetic environment. The
customer and/or user of the N-595 should assure that it is used in an
electromagnetic environment as described below.
IEC 60601-1-2
Immunity Test
Test Level
Power
3 A/m
frequency (50/
60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
Compliance
Electromagnetic
Level
Environment Guidance
3 A/m
If image distortion
occurs, it may be
necessary to position the
N-595 further from the
sources of power
frequency magnetic
fields or to install
magnetic shielding. The
power frequency
magnetic field should be
measured in the intended
installation location to
assure that it is
sufficiently low.