Replacement

N-595
6. Install new double-sided tape as shown in Figure 14 on page 63.
7. Slide the Display PCB into the grooves in the top case. Check to make
sure the Display PCB is firmly seated in the top case. Apply pressure
between the top case and the display PCB to make good contact with the
double-sided tape.
8. Connect the CCFL wire harness with two white wires to J5 of the User
Interface PCB.
9. Connect the Display PCB ribbon cable to J13 of the User Interface PCB.
Install the clip over the J5 connector.

Figure 14: Display PCB