Display/Alarms

N-595
Table 7 lists symptoms of problems relating to non-functioning displays and
audible tones or alarms, and recommended actions. If the action requires
replacement of a PCB or module, refer to Disassembly Guide on page 63.

Table 7: Display/Alarms Problems

Symptom
Display values are
missing or erratic.
Display pixels do not
light.
Alarm sounds for no
apparent reason.
Alarm does not sound.
Recommended Action
If the OXIMAX sensor is connected, replace the
pulse oximetry cable.
If the condition persists, replace the OXIMAX
sensor.
If the condition still persists, replace the User
Interface printed circuit board. See User Interface
PCB Removal/Replacement on page 76.
Check the connection between the User Interface
PCB and the Display PCB.
If the condition does not change, replace the
Display PCB. See Display PCB Removal/
Replacement on page 74.
If the condition still persists, replace the User
Interface PCB. See User Interface PCB Removal/
Replacement on page 76.
Moisture or spilled liquids can cause an alarm to
sound. Allow the monitor to dry thoroughly before
using.
If the condition persists, replace the User Interface
PCB. See User Interface PCB Removal/
Replacement on page 76.
Check alarm silence status.
Check speaker connection.
Replace the speaker as described in Alarm
Speaker Removal/Replacement on page 78.
If the condition persists, replace the User Interface
PCB. See User Interface PCB Removal/
Replacement on page 76.
Troubleshooting
55