Operational Setup

Alarms and Alarm
Silence
N-595
Operational set up procedures verify and setup the following parameters.
Alarms and Alarm Silence, page 15
Alarm Volume Control, page 17
Pulse Tone Volume Control, page 17
Nurse Call, page 18
Analog Output, page 19
Operating on Battery Power, page 20
1. Connect the DOC-10 monitor cable to the monitor sensor port.
2. Connect the DS-100 sensor to the DOC-10 cable and yourself.
3. Press the POWER ON/OFF button to turn the monitor on.
4. Press the SETUP softkey.
5. Press the VIEW softkey.
6. Press the PLETH softkey. Verify that the %SpO
SpO
and pulse rate.
2
7. Press the LIMITS softkey.
8. Press the SELECT softkey to select SpO
1.
Sensor Port
and BPM indicate your
2
lower alarm limit.
2
Performance Verification
15