Disassembly Guide
Prior to
Disassembly
Fuse
Replacement
52
1. Turn the N-595 off by pressing the POWER ON/OFF button.
2. Disconnect the monitor from the AC power source.
1. Complete the procedure in paragraph Prior to Disassembly on page 52.
2. Disconnect the power cord from the back of the monitor.
3. Remove the fuse drawer from the power module by pressing down on the
tab in the center and pulling out as shown in Figure 7 on page 52.

Figure 7: Fuse Removal

4. Put two new, 5 x 20-mm, slow blow, 0.5-amp, 250-volt fuses in the
drawer and reinsert the drawer in the power entry module.