Monitor
Disassembly
N-595
1. Complete the procedure in paragraph Prior to Disassembly on page 52.
2. Set the N-595 upside down, as shown in Figure 8 on page 53.

Figure 8: Corner Screws

3. Remove the monitor's four corner screws.
Caution: Observe ESD (electrostatic discharge) precautions when
disassembling and reassembling the N-595 and when handling any of the
components of the N-595.
4. Separate the monitor's top case from the bottom case of the monitor,
being careful not to stress the wire harnesses between the cases. Place the
two halves of the monitor on the table as shown in Figure 9 on page 54.
5. Disconnect the monitor's Power Supply harness from J16 on the User
Interface PCB.
Disassembly Guide