6.6.9 Total Session Status Window

The Total Session Status window indicates the Elapsed Test Time and the Transducer used.

6.6.10 Stimulus Rate

Select the appropriate stimulus rate (40Hz or 90Hz) for testing. To change during testing (when the patient
state changes), all test frequencies must have finished or be manually stopped.

6.7 Making an ASSR recording

6.7.1 Start & Stop
Start and Stop button are used for starting and stopping the measurement. Start button changes into a Stop
button when a recording is started. The Start button will be greyed out if no test stimuli are selected.

6.7.2 Pause

Pause becomes active after testing has started. It allows for pausing during a test. The stimulus will
continue to be presented, but no measurement is occurring. Click on Resume to resume testing.

6.7.3 Raw EEG Graphs

The Raw EEG graphs for Right and Left ears indicate the ongoing raw EEG. When the curves are black, the
EEG is within the set rejection level. When the curves turn red, the recorded response will be rejected as the
EEG exceeds the set rejection limits.
The rejection levels can be adjusted by clicking on the arrows between the two raw EEG curve graphs.
During testing the rejection level can only be adjusted within certain limits. To change the rejection level
outside these limits, the test must be stopped.