Right click on the curve's handle to show the differential Ipsi minus Contra curve.

3.2.13 Differential A minus B (A-B (N)) Curve

Right click on the curve's handle to show the differential A minus B curve.

3.2.14 Changing Display Filtering

Modify the display filters at any time during testing or when editing later to remove unwanted noise from the
recording.
Note, this will not affect the raw data recording (hardware filters) and the display filters can be
changed/disabled at any time!!

3.2.15 Recorded Curve Conditions

To display a curve's recording parameters, double click on the curve's handle and ensure Show conditions is
selected. The following information is displayed:
1. Measured – the number of sweeps recorded.
2. Rejected – the number of sweeps rejected.
3. Rejection – the rejection level used during testing.
4. Masking – indicates the masking level in dBSPL or off if not used.
5. Stim./Sec – the stimulus rate used during testing.