1-Point Manual Calibration - ABB AWT210 Operating Instruction

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C

1-point manual calibration

Performs a manual calibration (Offset adjustment) at a single
reference point.
1
Wait for stable reading
Monitor the process value and
continue (
) to the next step once
the value has stabilized.
2
Enter the new value
Enter the desired PV value by
pressing the
key to move the
cursor and the
keys to
change the value. When the new
value has been entered press the
key to continue.
3
Completion
Following a successful calibration
the calculated coefficients are
displayed.
Following an unsuccessful
calibration the reason for failure
is displayed.
2-point manual calibration
Performs a 2-point calibration using 2 pre-defined buffer
solutions.
1
– – –
1 Point Manual
PV
7.00 pH
Continue When Stable
Abort
Continue
– – –
1 Point Manual
2
PV
7.00 pH
0
07.12
New
Continue
Next
– – –
1 Point Manual
3
PV
7.12 pH
PV Slope
100.00%
PV Offset
8.94 mV
Exit
– – –
1 Point Manual
4
PV
7.14
pH
Calibration Failed
Offset Too High
Exit
5
6
Buffer temperature
The temperature of the buffer
solutions is displayed. The
temperature can be edited by
pressing the
key. When the
buffer temperature is correct press
the
key to continue.
Buffer 1 value
The value of the 1
st
buffer solution
is displayed. The value can be
edited by pressing the
key.
When the buffer value is correct
press the
key to continue.
Wait for stable reading – 1
buffer solution
Immerse the sensor in the buffer
solution, monitor the process value
and continue
to the next step
once the value has stabilized.
Buffer 2 value
The value of the 2
buffer solution
nd
is displayed. The value can be
edited by pressing the
key.
When the buffer value is correct
press the
key to continue.
Wait for stable reading –
2
buffer solution
nd
Immerse the sensor in the buffer
solution, monitor the process value
and continue
to the next step
once the value has stabilized.
Completion
Following a successful calibration
the calculated coefficients are
displayed.
Following an unsuccessful
calibration the reason for failure
is displayed.
– – –
2 Point Manual
Buffer Temperature
25.0 °C
Next
– – –
2 Point Manual
Buffer 1
4.00 pH
Next
st
– – –
2 Point Manual
PV
4.03 pH
Immerse In Buffer 1
Continue When Stable
Abort
Continue
– – –
2 Point Manual
Buffer 2
7 .00 pH
Next
– – –
2 Point Manual
PV
7.15 pH
Immerse In Buffer 2
Continue When Stable
Abort
Continue
– – –
2 Point Manual
PV
7.00 pH
PV Slope
98.75%
PV Offset
8.94 mV
Exit
– – –
2 Point Manual
PV
7.15 pH
Calibration Failed
Slope Too High
Exit