Tests and Adjustments
Performance Tests
Performance Tests
The only performance test applicable to your N5264A microwave PNA-X
measurement receiver is:
—
Frequency Accuracy Test
Function of the Test: To verify the frequency accuracy and range of the
analyzer's source output.
Specification Tested: Test Port Output–CW Accuracy
Equipment Used: A frequency counter, a test cable, and adapters.
Description of the Test:
1. Connect a frequency counter to the rear-panel 10 MHz REF OUT
If the Analyzer Fails this Test:
— If the 10 MHz reference is off by more than 10 Hz, perform the
Keysight N5264A Service Guide
Frequency Accuracy Test
connector.
Frequency Reference Adjustment" on page 3-6
"10 MHz
and then repeat this test.
3-5