Add 4-port capability for version 6 synthesizers (44 pages)
Summary of Contents for Keysight N5264A
Page 1
Keysight N5264A PNA-X Measurement Receiver Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. Service Guide...
DOCUMENT OR ANY INFORMATION reproduce, release, perform, CONTAINED HEREIN. SHOULD display, or disclose commercial KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT computer software or WITH WARRANTY TERMS commercial computer software COVERING THE MATERIAL IN THIS...
Page 5
Removing and Replacing the A1–A3 and Other Front Panel Subassemblies... . 7-9 Removing and Replacing the A10, A11, and A12 Boards (For analyzers with serial numbers prefixed MY/SG/US5144 and below.) ... . . 7-13 Keysight N5264A Service Guide...
Page 6
Recovering from Hard Disk Drive Problems ........8-7 Keysight N5264A Service Guide...
Keysight Measurement Receiver PNA-X Service Guide Safety and Regulatory Information Information in This Chapter This chapter provides safety information that will help protect you and your network analyzer. It also contains information that is required by various government regulatory agencies.
Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. Keysight N5264A Service Guide...
This product is designed for use in Installation Category II and Pollution Degree 2. Do not operate the analyzer with the outer cover removed for more than 30 minutes, as this could cause the analyzer to overheat which could result in costly damage. Keysight N5264A Service Guide...
Page 10
The measuring terminals on this instrument are designed to be used with external signals described in Measurement Category I, but NOT with external signals described in Categories II, III, and IV. The input of this instrument cannot be connected to the mains. Keysight N5264A Service Guide...
LINE switch (disconnecting device). There are no replaceable fuses in the mains input or within the power supply assembly. Keysight personnel: after calibration is completed, attach four “calibration void if seal broken” stickers to the PNA as shown in Figure 1-1.
Page 12
Safety and Regulatory Information General Safety Considerations Figure 1-1 Location of Calibration Stickers on PNA Keysight N5264A Service Guide...
3. Connect the other end of the cable to the test port and remove the short from the cable. Figure 1-2 shows a typical ESD protection setup using a grounded mat and “ESD Supplies” on page 6-33 wrist strap. Refer to for part numbers. Keysight N5264A Service Guide...
Page 14
Safety and Regulatory Information Electrostatic Discharge Protection Figure 1-2 ESD Protection Setup Keysight N5264A Service Guide...
This is a marking to indicate product compliance with the Canadian Interference-Causing ICES/NMB-001 Equipment Standard (ICES-001). Direct Current. IP 2 0 The instrument has been designed to meet the requirements of IP 2 0 for egress and operational environment. Keysight N5264A Service Guide...
If the battery on the A17 CPU board assembly needs to be disposed of, dispose of it in accordance with your country’s requirements. If required, you may return “Contacting the battery to Keysight Technologies for disposal. Refer to Keysight” on page 2-8 for assistance.
Keysight Support, Services, and The Internet address (URL) for on-line assistance. page 2-8 Assistance Service and support options available. Calibration options available. Important information about shipping your measurement receiver to Keysight for service or repair.
Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally. Electrical Maintenance “Review the Principles of Connector Care” on page 3-3. Refer to Keysight N5264A Service Guide...
Analyzer Options and Accessories Available Analyzer Options and Accessories Available To see a list of the options and accessories available for the network analyzers, including ordering information, refer to the Keysight PNA Family Microwave Network Analyzers Configuration Guide, available online at http://literature.cdn.keysight.com/litweb/pdf/5990-7745EN.pdf.
Refer to the previous section, Available” on page 2-3, for a complete description of each option upgrade. Table 2-1 Upgrades Available Upgrade Order Number Time Domain (Option 010) N5264AU-010 LO Source (Option 108) N5264AU-108 Fast CW Mode (Option 118) N5264AU-118 Keysight N5264A Service Guide...
Required Service Test Equipment (Cont’d) Critical Specifications Recommended Al ternate Model Equipment Model or Part or Part Number Number Calibration and Verification Kits 3.5 mm calibration kit Freq: 10 MHz to 26.5 GHz 85052B 85052D P,T,V Keysight N5264A Service Guide...
Page 22
1/4 inch and 5/16 inch open-end Thin profile 8710-0510 A, R wrench 5/16 inch, open-end torque wrench 1.1 and 2.4 N-m (10 and 21 T, R (metric equivalent is 8 mm) in-lb) settings (for semi-rigid cables) Keysight N5264A Service Guide...
Page 23
USB flash ROM drive None P, A, R a. Unless specified otherwise, equipment listed is required for all measurement receiver options. b. P = Performance tests, A = Adjustments, T = Troubleshooting, R = Repair, V = System verification Keysight N5264A Service Guide...
Service and Support Options The analyzer’s standard warranty period is three-years from the time of initial delivery. All repairs require the analyzer to be shipped to the nearest Keysight Technologies service center. Extended warranty periods can be purchased with the initial product purchase.
General Product Information Keysight Support, Services, and Assistance If you do not have access to the Internet, please contact your Keysight field engineer. In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number. With this information, the Keysight representative can determine whether your product is still within its warranty period.
Keysight Measurement Receiver PNA-X Service Guide Tests and Adjustments Information in This Chapter This chapter contains procedures to help you check, verify, and adjust your PNA. — The checks verify the operation of the assemblies in your analyzer. — The verification compares the operation of your analyzer to a gold standard.
To reduce the chance of electrostatic discharge, follow all of the recommendations outlined in “Electrostatic Discharge Protection” on page 1-7, for all of the procedures in this chapter. Keysight N5264A Service Guide...
— Use correct end of calibration block — Gage all connectors before first use Making Connections — Align connectors carefully Do Not — Apply bending force to connection — Make preliminary connection contact — Over tighten preliminary connection lightly Keysight N5264A Service Guide...
Page 29
Cleaning connectors with alcohol shall only be done with the instrument’s power cord removed, and in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument. Keysight N5264A Service Guide...
Tests and Adjustments Performance Tests Performance Tests The only performance test applicable to your N5264A microwave PNA-X measurement receiver is: Frequency Accuracy Test — Frequency Accuracy Test Function of the Test: To verify the frequency accuracy and range of the analyzer’s source output.
Adjustments are firmware-driven tests that are used to fine-tune your analyzer. The 10 MHz Frequency Reference Adjustment is applicable to the standard N5264A and the Synthesizer BW Adjustment and EE Default Adjustment are applicable for the N5264A Option 108. “10 MHz Frequency Reference Adjustment” on page 3-6 —...
, then , then , then click EE Default Adjustment. 2. On the dialog box, select Initialize rather than Adjust/Verify because an adjustment is typically unnecessary. 3. Follow the instructions and prompts as they are displayed. Keysight N5264A Service Guide...
Keysight Measurement Receiver PNA-X Service Guide Troubleshooting Information in This Chapter The information in this chapter helps you: — Identify the portion of the analyzer at fault. — Locate the specific troubleshooting procedure to identify the assembly or peripheral at fault.
This represents a “hardkey”, a key that is physically located on the instrument. Hardkey This represents a “softkey”, a key whose label is determined by the instrument Softkey firmware. Menu Item This represents an item in a drop-down or pop-up menu. Keysight N5264A Service Guide...
Operating the Analyzer With Covers Removed Operating the Analyzer With Covers Removed Do not operate the analyzer with the outer cover removed for more than 30 minutes, as this could cause the analyzer to overheat which could result in costly damage. Keysight N5264A Service Guide...
To reduce the chance of electrostatic discharge, follow all of the recommendations outlined in “Electrostatic Discharge Protection” on page 1-7, for all of the procedures in this chapter. Keysight N5264A Service Guide...
Step 2. Replace the faulty assembly and determine what adjustments are necessary. Refer to Chapter 7, “Repair and Replacement Procedures.” Step 3. Perform the necessary adjustments. Refer to Chapter 3, “Tests and Adjustments.” Step 4. Perform the necessary performance tests. Refer to Chapter 3, “Tests and Adjustments.” Keysight N5264A Service Guide...
Check the Basics A problem can often be solved by repeating the procedure you were following when the problem occurred. Before calling Keysight Technologies or returning the instrument for service, please perform the following checks: 1. Is there power at the mains receptacle? If not, correct this situation and proceed.
Page 39
Troubleshooting Getting Started with Troubleshooting Figure 4-1 Troubleshooting Organization Flowchart Go to “Power Up Troubleshooting” Go to “Front Panel Troubleshooting” “Rear Panel Go to Troubleshooting” Keysight N5264A Service Guide...
— If the line switch glows green and the fans are operating (audible), but the “Troubleshooting LCD Display Problems” on display remains dark, go to page 4-14. — If the instrument appears to abort the network analyzer measurement “Contacting Keysight” on interface process, contact Keysight. Refer to page 2-8. Keysight N5264A Service Guide...
ACOM as ground reference for analog supplies and the point marked DCOM as ground reference for digital supplies. Refer to Figure 4-2 for the power supply measurement points on the A15 Table 4-2 on page 4-11 midplane board. Refer to for the correct voltages. Keysight N5264A Service Guide...
Page 42
Failure to do so can result in personal injury or loss of life. Do not operate the analyzer with the outer cover removed for more than 30 minutes, as this could cause the analyzer to overheat which could result in costly damage. 4-10 Keysight N5264A Service Guide...
Page 43
If the missing supply voltages return to a “power on” condition after removal of an assembly, suspect that assembly as being defective. Remove the network analyzer assemblies in the order specified in the following Chapter 7 steps (refer to for removal instructions). Keysight N5264A Service Guide 4-11...
Page 44
You should be able to determine if the shorted condition has changed. Isolate the remaining three assemblies: 4-12 Keysight N5264A Service Guide...
Page 45
A14 system motherboard. THIS MUST BE DONE QUICKLY AS THE ANALYZER WILL RAPIDLY OVERHEAT WITHOUT THE COOLING EFFECT OF THE FANS. DO NOT PLUG IN THE POWER CORD UNTIL READY TO PERFORM MEASUREMENTS. Keysight N5264A Service Guide 4-13...
There are no front panel adjustments for intensity and contrast of the LCD. 1. If the display is dim, the A3 display assembly is defective. Refer to “Removing and Replacing the A1–A3 and Other Front Panel Subassemblies” on page 7-9. 4-14 Keysight N5264A Service Guide...
Page 47
4. If the front panel interface ribbon cable is properly connected, suspect that one or more of the following is defective: — inverter board (mounted on the display assembly) — A1 front panel interface board — A3 display assembly Keysight N5264A Service Guide 4-15...
Service More Front Panel Test Press UTILITY System , then , then , then A Front Panel Key Test Utility dialog box will be displayed, as shown in Figure 4-4. 4-16 Keysight N5264A Service Guide...
Page 50
3. If the movement of numbers is not smooth or no numbers appear at all, suspect a faulty A1 front panel interface board. To replace the A1 front panel interface board, refer to “Removing the A1 Front Panel Interface Board and Keypad Assembly” on page 7-9. 4-18 Keysight N5264A Service Guide...
USB port. — Wait 15 seconds for the analyzer to verify the device connection, and then check the operation of the USB device. — If the device performs correctly, the USB board is functioning properly. Keysight N5264A Service Guide 4-19...
Checking the Operation of the Key Pad Commands To verify the key pad functionality, refer to “Front Panel Keypad and RPG Test” on page 4-16. Checking the Display To verify the display functionality, refer to “A3 Display Test” on page 4-19. 4-20 Keysight N5264A Service Guide...
, then Configure , then SICL/GPIB... SICL/GPIB/SCPI dialog box is displayed. 3. In the GPIB block, click System Controller to establish the analyzer as a controller. Wait for the analyzer to configure, and then click OK. Keysight N5264A Service Guide 4-21...
Problems with the Local Area Network (LAN) can be difficult to solve. Software and protocol problems can make it difficult to determine whether the analyzer's hardware is working properly, or if there is a problem with the LAN or cabling. 4-22 Keysight N5264A Service Guide...
Page 55
Windows Taskbar menu at the bottom of the display. b. On the Windows Taskbar menu, click Start, point to Programs, Accessories, and then click Command Prompt. c. The command prompt window is displayed. Keysight N5264A Service Guide 4-23...
Page 56
Other software may behave somewhat differently, but basically the same. 1. The letters x represent the IP address of the other device on the network. 2. The letters x represent the IP address of the analyzer. 4-24 Keysight N5264A Service Guide...
Page 57
A crossover cable can be made from a standard LAN cable by connecting pin 1 from each connector to pin 3 of the other connector, and pin 2 from each connector to pin 6 of the other connector. Keysight N5264A Service Guide 4-25...
Page 58
2 on the other end of the cable. 3. Insulate all exposed wires so that they cannot short together. 4. Label this as a crossover cable so that it cannot be confused with a standard cable. 4-26 Keysight N5264A Service Guide...
Page 59
Troubleshooting Rear Panel Troubleshooting Figure 4-5 Construction of a Crossover Cable Keysight N5264A Service Guide 4-27...
Page 61
REAR PANEL FRONT PANEL N5264A Overall Block Diagram INTERCONNECTS INTERCONNECTS (Includes Options 700 - Standard; 108 - LO Source Service Guide: N5264-90001 USB x 4 DISPLAY USB BD A20 IF MULTIPLEXER A12 SIGNAL PROCESSING Main IF A ADC MODULE (SPAM) 10.7 MHz...
Theory of Operation Information in This Chapter This chapter provides a general description of the operating theory of the N5264A PNA-X measurement receiver. — Theory of operation is explained to the assembly level only. — Component-level circuit theory is not provided.
The major functional groups are: — Synthesized LO Group (Option 108) IF Multiplexer Group — — Digital Processor and Digital Control Group Power Supply Group — Keysight N5264A Service Guide...
Page 65
“Digital Processing and Digital Control Group Operation” on Refer to page 5-14. Power Supply Group The power supply functional group provides power for the other assemblies in the instrument. “Power Supply Group Operation” on page 5-19. Refer to Keysight N5264A Service Guide...
50 MHz that is sent to the A11 13.5 GHz LO synthesizer board. The A11 13.5 GHz synthesizer board produces an LO signal that is sent through the A21 multiplier/amplifier 26.5 (HMA26.5) to the rear panel LO OUT connector. Keysight N5264A Service Guide...
Page 67
Theory of Operation Synthesized LO Group Operation (Option 108) Figure 5-2 Synthesized LO Group (Option 108) Keysight N5264A Service Guide...
10 MHz reference signal for the A12 signal processing ADC module (SPAM) board. The VCO output is also divided by two to produce a 50 MHz reference signal for the A11 13.5 GHz LO synthesizer board. Keysight N5264A Service Guide...
MHz to 26.508 GHz that is sent to the rear panel LO OUT connector. A19 Test Set Motherboard The A19 test set motherboard serves these functions: — to act as an interface between the A17 CPU board and the auxiliary rear panel interconnects. Keysight N5264A Service Guide...
Page 70
TTL in, low level (10 s min) aborts sweep Interrupt In +22 V +22 Vdc, 100 mA max. 15–16 SEL1–2 TTL out, test set select bits 1-2, tied to 0 V AD11 TTL I/O, address and latched data Keysight N5264A Service Guide...
Page 71
TTL out, low = input mode, high = output mode TTL out, active low data write strobe (1 s min) Output Strobe Write Strobe No connect Not used Pass Fail TTL out, latched, indicates pass fail (programmable polarity) Keysight N5264A Service Guide...
Page 72
Programmable ±10 V @ 100 mA out Nominally 0 ohms 2.44 mV typical resolution 1 MHz BW ACOM System ground GndSense Ground sense for Analog In and Analog Out Connected with 51.1 ohms to ACOM 5-10 Keysight N5264A Service Guide...
Page 73
Analog input: ±10 V @ 1.22 mV typical resolution Rin > 1 M-ohm BW 1 MHz ADC conversion time < 1 us typical Power Button Open collector input Active low replicates power button key press. Keysight N5264A Service Guide 5-11...
These inputs provide the five IF input signals that are sent to the A12 SPAM board for digital processing. A20 IF Multiplexer Board The A20 IF multiplexer board routes the five rear panel IF input signals to the A12 SPAM board. 5-12 Keysight N5264A Service Guide...
Page 75
Theory of Operation IF Multiplexer Group Operation Figure 5-3 IF Multiplexer Group Keysight N5264A Service Guide 5-13...
— — A2 USB Board A3 Display Assembly — — Keypad Assembly — A14 system motherboard Data Acquisition and Processing Subgroup — A12 SPAM Board — — A17 CPU Board A51 Solid State Drive — 5-14 Keysight N5264A Service Guide...
Page 77
Theory of Operation Digital Processing and Digital Control Group Operation Figure 5-4 Digital Processing and Digital Control Group Keysight N5264A Service Guide 5-15...
The video data received from the A17 CPU board includes the following: • digital TTL horizontal sync • digital TTL red video • blanking • digital TTL vertical sync • digital TTL green video • data clock • digital TTL blue video 5-16 Keysight N5264A Service Guide...
In addition, when the analyzer is in the system controller mode, the CPU controls peripheral devices through the peripheral port interfaces. Front panel settings are stored in SRAM, with a battery providing at least five years of backup storage when external power is off. Keysight N5264A Service Guide 5-17...
Page 80
(IDE) data storage device which is connected directly to, and physically mounted within the enclosure of, the A17 CPU board. The full operating system and firmware for the network analyzer is stored on the A51 hard disk drive. 5-18 Keysight N5264A Service Guide...
The +15 V standby supply remains on continuously whenever the power supply is plugged in. This supply is used to provide power to front panel LEDs and CPU components when the analyzer is turned off. Keysight N5264A Service Guide 5-19...
Service Guide Replaceable Parts Information in This Chapter This chapter: — identifies the replaceable parts for the Keysight PNA-X measurement receiver. — includes several tables and illustrations to assist you in identifying the correct part for your measurement receiver. — contains ordering information for new assemblies and rebuilt-exchange assemblies.
Ordering Information To order a part listed in the replaceable parts lists: — include the part number — indicate the quantity required — Contact Keysight Technologies for instructions on where to send the order. Refer to “Contacting Keysight” on page 2-8.
Adjustments.” Step 6. Perform the necessary performance tests. Refer to Chapter 3, “Tests and Adjustments.” Step 7. Keysight personnel: see Figure 1-1 on page 1-6 to review where the calibration stickers should be placed on the PNA. Keysight N5264A Service Guide...
Replaceable Parts Replaceable Parts Listings Replaceable Parts Listings This section contains the replacement part numbers and their descriptions for your Keysight microwave PNA. You can find the locations of replaceable parts in this section: Table 6-1, or — listed by reference designator in Table 6-2.
Page 86
Memory Modules on page 6-32 — Upgrades Orderable by Model Number on page 6-32 — USB Accessories on page 6-33 — Rack Mount Kits and Handle Kits on page 6-33 — Touch-up Paint on page 6-33 — Keysight N5264A Service Guide...
Power switch overlay ➆ W1312-40017 Front (RPG) knob ⑧ N5264-80002 Nameplate Not shown 5023-1399 Front handle a. For part numbers of rack mount kits and handle kits, refer to “Rack Mount Kits and Handle Kits” on page 6-33. Keysight N5264A Service Guide...
J1 ➀ N5242-40001 Keypad assembly ➁ 0515-0430 Machine screw, M3.0 x 6 pan head (9 to attach front panel interface board to front frame and 3 to attach display cable to display hold down bracket.) Keysight N5264A Service Guide...
Page 89
2090-1036 (LED) LCD display 2090-0883 (Tube) critical footnote Display assembly (continued from previous page) N5242-20121 Spacers (For display hold down bracket standoffs.) For use with LCD (Tube) display 2090-0883 only. critical footnote Keysight N5264A Service Guide...
Page 90
In March 2013, Keysight discontinued using a display with florescent tube backlighting, and began using a display with LED backlighting. Associated parts are indicated in this table with either “Tube” or “LED.” If you replace an old part number <designated “Tube”>...
Page 91
Replaceable Parts Replaceable Parts Listings Figure 6-2 Front Panel Assembly, Back Side, All Options Figure 6-3 A3 Display Assembly with Old LCD Display 2090-0883 6-10 Keysight N5264A Service Guide...
Replaceable Parts Listings Figure 6-4 A3 Display Assembly with New LCD Display 2090-1036 Top Assemblies, All Options When replacing an old assembly, install an assembly with either the same part number or the new part number. Keysight N5264A Service Guide 6-11...
Page 93
Replaceable Parts Replaceable Parts Listings In February 2012, the N5264A analyzer underwent significant hardware changes. These changes included a redesigned Signal Processing ADC Module (SPAM) board, Frequency Reference board, left side inner bracket, and right side inner bracket. It is very important that this redesigned hardware be used only with analyzer serial numbers prefixed MY/SG/US5204 and above.
Page 94
Windows XP, and you do not want to upgrade to Windows 7, it is highly recommended that you upgrade your firmware to 9.90.17 or above. Please be aware that Keysight cannot guarantee 100% compatibility with Windows XP and or older versions of firmware <10.45.xx. To upgrade your firmware go to http://na.support.key-...
Page 95
Replaceable Parts Replaceable Parts Listings Figure 6-5 Top Assemblies, All Options 6-14 Keysight N5264A Service Guide...
Page 96
Replaceable Parts Replaceable Parts Listings This page is intentionally left blank. Keysight N5264A Service Guide 6-15...
Replaceable Parts Listings Bottom Assemblies, All Options Reference Part Number Description Designator N5240-60163 Test set motherboard Was N5245-60003 or N5240-60043 N5240-60062 IF multiplexer board Was N5240-60045 5087-7765 LO Multiplier/amplifier 26.5 (HMA26.5) (Option 108) Was 5087-7711 6-18 Keysight N5264A Service Guide...
Page 100
Replaceable Parts Replaceable Parts Listings Figure 6-7 Bottom Assemblies, All Options Keysight N5264A Service Guide 6-19...
A20 IF multiplexer board J1 ➃ 8121-0834 Rear-panel HANDLER I/O to A19 test set motherboard J400 a. SR = semirigid coaxial cable; F = flexible coaxial cable; nR = n wires in a ribbon (flat) cable 6-20 Keysight N5264A Service Guide...
Page 102
Replaceable Parts Replaceable Parts Listings Figure 6-8 Bottom Cables, All Options Keysight N5264A Service Guide 6-21...
Power supply bracket ➇ 0515-0375 Machine screw, M3.0 x 16, pan head (To attach midplane board to midplane bracket.) ➈ 0400-0353 Midplane board grommets Figure 6-9 Top Hardware and Miscellaneous Parts, Side View, All Options 6-22 Keysight N5264A Service Guide...
Page 104
Replaceable Parts Replaceable Parts Listings Figure 6-10 Top Hardware and Miscellaneous Parts, Top View, All Options Keysight N5264A Service Guide 6-23...
➆ N5221-00006 Test set front sub panel N5242-00018 ➇ 0515-1227 Machine screw M3.0 x 6 flat head (to attach test set front sub panel to front of side brackets.) ➈ N5221-00005 Test set deck N5242-00002 6-24 Keysight N5264A Service Guide...
Page 106
Replaceable Parts Replaceable Parts Listings Figure 6-11 Bottom Hardware and Miscellaneous Parts, All Options Keysight N5264A Service Guide 6-25...
➆ All prefixes N5242-40002 PC board guides ➇ All prefixes N5242-00016 Side bracket (left) a. In this table, the two letters that indicate the PNA manufacturing location have been removed from each serial number prefix. 6-26 Keysight N5264A Service Guide...
Page 108
Replaceable Parts Replaceable Parts Listings Figure 6-12 Internal Hardware and Miscellaneous Parts, All Options Keysight N5264A Service Guide 6-27...
Hole plug (When instrument is rack mounted.) ➇ 0515-0372 Machine screw, M3.0 x 8, pan head (To attach inner cover.) ➈ N5221-00001 Inner cover (retaining shield) Was N5264-00003 ➉ 0515-1227 Machine screw, M3.0 x 6, flat head (To attach inner cover.) 6-28 Keysight N5264A Service Guide...
Page 110
Replaceable Parts Replaceable Parts Listings Figure 6-13 External Hardware and Miscellaneous Parts, All Options Keysight N5264A Service Guide 6-29...
Installation and Quick Start Guide (for all PNA series analyzers) (Cannot be ordered. Part number E8356-90001 is for reference only. Must be printed from the Keysight Web site.) Service Guide. (Not available in printed form. Part number is for reference only. Must be printed N5264-90001 from the Keysight Web site.)
Page 114
N5232AU-1CP or N5239AU-1CP Option 1CP includes the following separately orderable items: Rack mount kit (rack mount flanges and hardware) 5063-9237 Rack mount rail set E3663AC 5063-9230 Front handle kit (two classic handles and hardware) Touch-up Paint Keysight N5264A Service Guide 6-33...
Page 115
Parchment white (for use on rack mount flanges, rack support flanges, and front 6010-1148 panels) a. For rack mount use, you must replace factory installed ruggedized handles (thick aluminum, no trim) with classic handles (thin aluminum with plastic trim), included with Option 1CP. 6-34 Keysight N5264A Service Guide...
PNA-X Service Guide Repair and Replacement Procedures Information in This Chapter This chapter contains procedures for removing and replacing the major assemblies of your Keysight Technologies PNA-X measurement receiver. Chapter Seven at-a-Glance Section Title Summary of Content Start Page Personal Safety Warnings Warnings and cautions pertaining to personal safety.
The front panel switch is only a standby switch and is not a LINE switch (disconnecting device). Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to manufacturer’s instructions. Keysight N5264A Service Guide...
To reduce the chance of electrostatic discharge, follow all of the recommendations outlined in “Electrostatic Discharge Protection” on page 1-7, for all of the procedures in this chapter. Keysight N5264A Service Guide...
Ribbon cables and wires must never be positioned on top of hardware. 3. Reverse the order of the removal procedures above. Keysight N5264A Service Guide...
Page 121
Repair and Replacement Procedures Removing the Covers Figure 7-1 Outer and Inner Cover Removal Keysight N5264A Service Guide...
) from the A1 front panel interface board. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. Keysight N5264A Service Guide...
Page 123
Repair and Replacement Procedures Removing and Replacing the Front Panel Assembly Figure 7-2 Front Panel Assembly Removal Keysight N5264A Service Guide...
) and remove the inverter board. Removing the Touchscreen Controller Board ④ 1. Disconnect the touchscreen controller board cable (item ) from the touchscreen controller board and front panel interface board. 2. Disconnect the flat flex cable from the touchscreen. Keysight N5264A Service Guide...
Page 125
A3 display assembly from the front panel assembly. 3. The touch screen can now be removed from the front panel assembly. Note the orientation of the touch screen in the front panel assembly for installation of the new touch screen. 7-10 Keysight N5264A Service Guide...
Page 126
(item ) that attach it to ⑧ the LCD display hold down bracket (item ). Note the location of the four ⑩ spacers (item ) before separating the LCD display from the hold down bracket. Keysight N5264A Service Guide 7-11...
Page 127
A3 Display Assembly and Touch Screen Removal-2 Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-12 Keysight N5264A Service Guide...
Removing and Replacing the A10, A11, and A12 Boards (For analyzers with serial numbers prefixed MY/SG/US5144 and below.) In February 2012, the N5264A analyzer underwent significant hardware changes. These changes included a redesigned Signal Processing ADC Module (SPAM) board, left side inner bracket, and right side inner bracket.
Page 129
Repair and Replacement Procedures Removing and Replacing the A10, A11, and A12 Boards (For analyzers with serial numbers prefixed MY/SG/US5144 and below.) Figure 7-6 A10, A11, and A12 Boards Removal 7-14 Keysight N5264A Service Guide...
Removing and Replacing the A10, A11, and A12 Boards (For analyzers with serial numbers prefixed MY/SG/US5204 and above.) In February 2012, the N5264A analyzer underwent significant hardware changes. These changes included a redesigned Signal Processing ADC Module (SPAM) board, left side inner bracket, and right side inner bracket.
Page 131
MY/SG/US5204 and above.) 3. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. Figure 7-7 A10, A11, and A12 Boards Removal 7-16 Keysight N5264A Service Guide...
) that secure the A14 system motherboard to the chassis. 12.Slide the A14 system motherboard toward the rear of the measurement ⑤ receiver to release it from the 15 keyhole standoffs (item ) on the chassis. Keysight N5264A Service Guide 7-17...
Page 133
Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-18 Keysight N5264A Service Guide...
Page 134
Repair and Replacement Procedures Removing and Replacing the A14 System Motherboard Figure 7-8 A14 System Motherboard Removal Keysight N5264A Service Guide 7-19...
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-20 Keysight N5264A Service Guide...
Page 136
Repair and Replacement Procedures Removing and Replacing the A15 Midplane Board Figure 7-9 A15 Midplane Board Removal Keysight N5264A Service Guide 7-21...
5. Slide the A16 power supply assembly out the rear of the measurement receiver. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-22 Keysight N5264A Service Guide...
Page 138
Repair and Replacement Procedures Removing and Replacing the A16 Power Supply Assembly Figure 7-10 A16 Power Supply Assembly Removal Keysight N5264A Service Guide 7-23...
(SSD)” on page 7-34. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-24 Keysight N5264A Service Guide...
Page 140
Repair and Replacement Procedures Removing and Replacing the A17 CPU Board Assembly Figure 7-11 A17 CPU Board Assembly Removal Keysight N5264A Service Guide 7-25...
5. Slide the A18 GPIB board out the rear of the measurement receiver. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-26 Keysight N5264A Service Guide...
Page 142
Repair and Replacement Procedures Removing and Replacing the A18 GPIB Board Figure 7-12 A18 GPIB Board Removal Keysight N5264A Service Guide 7-27...
Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-28 Keysight N5264A Service Guide...
Page 144
Repair and Replacement Procedures Removing and Replacing the A19 Test Set Motherboard Figure 7-13 A19 Test Set Motherboard Removal Keysight N5264A Service Guide 7-29...
Torque rear panel RF connector nuts to 21 in-lbs and PULSE I/O connector screws to 6 in-lbs. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-30 Keysight N5264A Service Guide...
Page 146
Repair and Replacement Procedures Removing and Replacing the A20 IF Multiplexer Board Figure 7-14 A20 IF Multiplexer Board Removal Keysight N5264A Service Guide 7-31...
1. Reverse the order of the removal procedure. Torque all RF cable connectors to 10 in-lbs. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-32 Keysight N5264A Service Guide...
Page 148
Repair and Replacement Procedures Removing and Replacing the A21 HMA26.5 Figure 7-15 A21 HMA26.5 Removal Keysight N5264A Service Guide 7-33...
Copy Unique Files from the Hard Disk If installing an SSD for Windows XP: if the user has loaded unique calibration kit information, navigate to C:\Program Files\Keysight\Network Analyzer and copy USER_CALKITFILE to a USB flash memory drive. Also copy any personal user files that you wish to preserve.
Page 150
Install Backup Files onto the New SSD The files that were previously saved onto a USB flash memory drive must now be installed onto the new SSD. The network analyzer must be powered up and operating. Keysight N5264A Service Guide 7-35...
9. Slide the rear panel over the cable connectors and off of the measurement receiver. Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-36 Keysight N5264A Service Guide...
Page 152
Repair and Replacement Procedures Removing and Replacing the Rear Panel Figure 7-17 Rear Panel Removal Keysight N5264A Service Guide 7-37...
Replacement Procedure 1. Reverse the order of the removal procedure. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-38 Keysight N5264A Service Guide...
Page 154
Repair and Replacement Procedures Removing and Replacing the Fans Figure 7-18 B1 Fan Removal Keysight N5264A Service Guide 7-39...
1. Reverse the order of the removal procedure following all instructions included with the new battery. 2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal procedure. Refer to Table 7-2 on page 7-42. 7-40 Keysight N5264A Service Guide...
Page 156
Repair and Replacement Procedures Removing and Replacing the Lithium Battery Figure 7-19 Lithium Battery Removal Keysight N5264A Service Guide 7-41...
A19 test set motherboard Reinstall the serial number. (Refer to None needed “Installing or Changing a Serial Number” Chapter Re-enable all hardware options. (Refer “Option Entitlement Certificate” Chapter A20 IF multiplexer board None needed None needed 7-42 Keysight N5264A Service Guide...
Page 158
A21 HMA26.5 (Option 108) EE Default Adjustment (Option None needed 108) A51 hard disk drive None needed Read and write to the drive B1 fan No adjustment needed Check for fan operation Battery No adjustment needed None Keysight N5264A Service Guide 7-43...
Keysight Measurement Receiver PNA-X Service Guide General Purpose Maintenance Procedures Information in This Chapter Chapter Eight at-a-Glance Section Title Summary of Content Start Page Option Enable Utility How to use the option enable utility to: page 8-2 — enable options that have been added to your measurement receiver, —...
Using the current Host ID will cause a license to be generated that will not work with the instrument. To determine the new Host ID, use the utility at the below listed web site with the new model number. http://na.support.keysight.com/pna/upgrades.html Keysight N5264A Service Guide...
Some applications require a license key that is provided by Keysight. If you do not have the required license key, contact Keysight for assistance. Refer to “Contacting Keysight”...
Page 162
, then , then System Help About NA... b. Press UTILITY System , then , then 6. If the options are still not listed correctly, contact Keysight for assistance. Refer to “Contacting Keysight” on page 2-8. Keysight N5264A Service Guide...
Change. 3. If an error is made in entering the serial number, obtain a clear code from Keysight, enter the clear code in the space provided and click CLEAR. The correct serial number can then be entered. To change an incorrect serial number, a clear-code password is required.
If your measurement receiver is connected to the Internet, there are two methods available for checking the availability of, and downloading, new firmware: — Download directly from: http://www.keysight.com/find/pna. (Select your measurement receiver’s model number in this web site to view available upgrades.) —...
Recovering from Hard Disk Drive Problems If you suspect that you have a hard disk drive problem, go to the “Hard Drive Recovery” link on the Keysight PNA Series: Service & Support Home Page on the Internet. The URL for the Keysight PNA Series: Service & Support Home Page is: http://na.support.keysight.com/pna/...
Page 166
A2 USB board illustrated , 7-14, 7-16 adjustments illustrated , 6-10, 6-11 procedure , 7-13, 7-15 EE default part number A14 system motherboard frequency removal and replacement illustrated 6-14 at 10 MHz illustrated 7-10 operation 5-16 synthesizer bandwidth Keysight N5264A Service Guide...
Page 167
5-16 part numbers 6-32 part number 6-28 rear panel , 5-7, 5-8, 5-18 required for servicing upgrades Display 5-18 warmup time EXT 10 MHz REF IN bottom assembly, assemblies location of part numbers bottom replacement assembly, Keysight N5264A Service Guide...
Page 168
See illustrated description 6-14 functional operation enabling digital control part number 6-11 options group removal entitlement certificate, digital processing replacement option group environment, operating 13, 7-14, 7-15, 7-16 IF multiplexer front frame specifications multiplexer illustrated verification Keysight N5264A Service Guide...
Page 169
6-11 6-14 illustrated 7-33 parts list part number 6-11 procedure 7-32 LED board removal illustrated , 6-19, replacement 6-25 part number , 6-18, 6-24 7-27 lower dress panel cables illustrated part numbers 6-32 data storage 5-18 Keysight N5264A Service Guide...
Page 170
, 6-10, 6-11 illustrated , 6-10, 6-11 parts list er board part number LED board CPU board 5-17 removal and replacement illustrated , 6-19, 6-25 data acquisition illustrated 7-10 part number , 6-18, 6-24 subgroup 5-17 procedure Keysight N5264A Service Guide...
Page 171
5-17 required for servicing solid state drive illustrated 5-18 , 6-22, 6-23 monitor, external 4-15 SPAM board miscellaneous 5-17 motherboard, system, See system motherboard 5-16 part numbers 6-32 test set motherboard required option Keysight N5264A Service Guide...
Page 172
ESD , 1-7, 3-2, 4-3, 4-4, screws replaceable cables, ribbon, for bottom assemblies pulse I/O adapter, N1966A See cables, illustrated 6-25 PWR I/O 6-16 ribbon part numbers 6-24 connector Keysight N5264A Service Guide...
Page 173
7-34 operation removal and replacement data storage 5-18 synthesizer band width illustrated , 7-11, 7-12 illustrated 6-14 adjustment procedure operation 7-10 5-18 system touch-up part number 6-11 adjustments paint removal and replacement EE default Keysight N5264A Service Guide...
Page 174
4-15 order numbers 6-32 port part numbers rear panel 6-32 to measurement receiver connector 5-18 accessories part numbers 6-33 board warm up front panel, analyzer warning illustrated , 6-10, Keysight N5264A Service Guide...