SMC Networks HRSH090-A Operation Manual page 136

Thermo chiller hrsh090 series
Table of Contents

Advertisement

HRX-OM-R026
Chapter 5
Display and setting of various functions
26.
Select responce delay time (simple communication protocol) from the table below with
[▲] key or [▼] key, and confirm by pressing "SEL".
Set value
to
Communication range (simple communication protocol) Setting and checking
27.
Press the [SEL] key once.
The set screen of communication range (simple communication protocol) is displayed
on the digital display.
PV
SV
28.
Select communication range (simple communication protocol) from the table below with
[▲] key or [▼] key, and confirm by pressing "SEL".
Set value
Contact input signal 1 Setting and checking
29.
Press the [SEL] key once.
The set screen of contact input signal 1 is displayed on the digital display.
PV
SV
5.21 Communication function
5-74
Table 5.21-14 List of set value
Explanation
Setting/checking are not available unless the
serial
protocol
setting
communication protocol.
Setting of response delay time.
Set range is 0 to 250m sec.
Table 5.21-15 List of set value
Explanation
Setting/checking are not available unless the
serial
protocol
setting
communication protocol.
Only reading is available
Reading and writing are available
Initial value
(Default setting)
is
simple
Initial value
(Default setting)
is
simple