6.4 Instantaneous Printing of Measurements
Following a measurement or after accessing values in MR mode, use the PRINT
function to print the measurement results.
Press the PRINT button to print the following:
• 1 set of measurements (V /R /DAR /PI/DD /date /time) for normal tests.
• The R(t) values if the "time-controlled test" has been used.
To stop printing, move the position of the selector switch.
Depending on the function used, the following models are obtained:
Insulation Measurement
AEMC
Instrument number: 000 001
INSULATION RESISTANCE TEST
OBJECT: 01 ..................................... TEST: 01 (only printed in MR mode)
Description: ....................................................
........................................................................
Date: ............................................03.31.2001
Start time: ............................................14 :55
Test duration: ......................... 15 min. 30 sec
Temperature: ..............................°C ......... °F
Relative humidity: ...................................... %
Test voltage: ....................................... 1000V
Insulation resistance (R): ............... 385GOhm
DAR (R 1'/R 30"): .................................1.234
PI (R 10'/R 1"): ......................................2.345
Comments: ....................................................
........................................................................
Date of next test: ............................../..../.......
After a 'time-controlled test' other results are printed (interval samples):
Time
0 0 : 30
0 1 : 00
0 1 : 30
etc...
A space for the operator's signature is available at the end of the printout.
Megohmmeter Models 5050/5060
0.561.8187
NOTE: PRINT only works when the test is stopped by pressing the START/STOP
button and the value is displayed on the LCD.
Instruments Model 5060
®
Resistance
35.94 GOhm
42.00 GOhm
43.50 GOhm